PizeNews

Pize adalah asisten pemrograman AI untuk komputasi ilmiah dan analisis statistik.

Akun

Sumber dengan panjang gelombang yang dapat ditala

ORPHEUS-VIS

OPA cahaya tampak berpita lebar

A broad-bandwidth visible OPA that delivers ultrashort VIS pulses for spectroscopy and imaging.

Fitur teknis

  • Broad-bandwidth visible output
  • Ultrashort VIS pulses
  • Yb-pumped ORPHEUS platform
  • Computer-controlled tuning
  • Spectroscopy and imaging

Spesifikasi teknis

ConfigurationStandardWide tuning range
Main output
Rentang penyetelan450 – 600 nm320 – 900 nm
Maximum pump power20 W20 W
Energi pulsa pompa200 – 1000 µJ200 – 1000 µJ
Durasi pulsa< 50 fs @ 500 – 600 nm< 50 fs @ 500 – 600 nm < 55 fs @ 800 – 900 nm < 70 fs @ 650 – 800 nm
Conversion efficiency 1)1.5 % @ 500 nm> 1.5% @ 500 nm > 5.0% @ 660 nm > 0.5% @ 350 nm
Spectral bandwidth 2)200 – 700 cm⁻¹200 – 700 cm⁻¹
Long-term power stability, 8 h 3)< 2% @ 500 nm< 2% @ 500 nm
Optional extension (UV)
Rentang penyetelan250 – 300 nm250 – 300 nm
Conversion efficiency 1)> 0.15% @ 280 nm> 0.15% @ 280 nm
Spectral bandwidth 2)200 – 600 cm⁻¹200 – 600 cm⁻¹
Pump laser requirements
ConfigurationCARBIDE or PHAROSCARBIDE or PHAROS
Panjang gelombang pusat1030 ± 10 nm1030 ± 10 nm
Maximum pump power20 W20 W
Laju pengulanganSingle-shot – 100 kHzSingle-shot – 100 kHz
Energi pulsa pompa200 – 1000 μJ200 – 1000 μJ
Pump pulse duration 4)200 – 350 fs200 – 350 fs
Environmental & utility requirements
Operating temperature 5)19 – 25 ºC (air conditioning recommended)19 – 25 ºC (air conditioning recommended)
Relative humidity 5)20 – 70% (non-condensing)20 – 70% (non-condensing)
Electrical requirements100 – 240 V AC, 1.4 A; 50 – 60 Hz100 – 240 V AC, 1.4 A; 50 – 60 Hz
Daya terukur120 W120 W
Power consumptionStandby: 10 W, Max during wavelength tuning: 100 WStandby: 10 W, Max during wavelength tuning: 100 W
  1. Specified as a percentage of pump power.
  2. Full width at half maximum (FWHM).
  3. Expressed as normalized root mean squared deviation (NRMSD).
  4. Full width at half maximum (FWHM), assuming a Gaussian pulse shape.
  5. Specifications are guaranteed for a maximum temperature variation of ± 1 ºC and humidity variation of ± 10%.