PizeNews

Pize es un asistente de programación con IA para cómputo científico y análisis estadístico.

Cuenta

Fuentes sintonizables

ORPHEUS-MIR

OPA de infrarrojo medio de banda ancha

A broad-bandwidth mid-infrared OPA for vibrational spectroscopy and MIR nonlinear optics.

Características principales

  • Broad-bandwidth MIR pulses
  • Yb-pumped ORPHEUS platform
  • Computer-controlled tuning
  • Built for vibrational spectroscopy
  • Pairs with CARBIDE or PHAROS

Especificaciones técnicas

Mode of operationNo colinealCollinear 1)
Main output (2500 – 10 000 nm)
Rango de sintonía2500 – 4000 nm (Signal) 4000 – 10 000 nm (Idler)2500 – 4500 nm (Signal) 2) 4500 – 10 000 nm (Idler)
Maximum pump power80 W80 W
Energía de pulso de bomba200 µJ – 3 mJ200 µJ – 3 mJ
Tasa de repetición máxima100 kHz100 kHz
Duración de pulso< 100 fs< 400 fs (< 100 fs with dispersion compensation) 1)
Conversion efficiency 3)> 1.2% @ 3000 nm > 1.0% @ 3500 nm > 0.6% @ 5000 nm > 0.3% @ 9000 nm> 1.2% @ 3000 nm > 1.0% @ 3500 nm > 0.6% @ 5000 nm > 0.3% @ 9000 nm
Spectral bandwidth 4)> 300 cm⁻¹ @ 3000 – 4000 nm > 200 cm⁻¹ @ 4000 – 10 000 nm> 300 cm⁻¹ @ 3000 – 4000 nm > 200 cm⁻¹ @ 4000 – 10 000 nm
Long-term power stability, 8 h 5)< 2% @ 5000 nm< 2% @ 5000 nm
Pulse-to-pulse energy stability, 1 min 5)< 2% @ 5000 nm< 2% @ 5000 nm
Auxiliary output (2000 nm)
Output wavelength 6)2000 ± 100 nm2000 ± 100 nm
Duración de pulso< 50 fs< 50 fs
Conversion efficiency 3)> 8%> 8%
Spectral bandwidth> 350 cm⁻¹> 350 cm⁻¹
Wavelength extension (10 000 – 15 000 nm)
Rango de sintonía10 000 – 15 000 nmn/a
Conversion efficiency 3)> 0.2% @ 12 000 nm
Spectral bandwidth> 100 cm⁻¹
Pump laser requirements
ConfigurationCARBIDE or PHAROSCARBIDE or PHAROS
Longitud de onda central1030 ± 10 nm1030 ± 10 nm
Maximum pump power80 W80 W
Tasa de repeticiónSingle-shot – 100 kHzSingle-shot – 100 kHz
Energía de pulso de bomba200 µJ – 3 mJ200 µJ – 3 mJ
Pump pulse duration 7)180 – 500 fs180 – 500 fs
Environmental & utility requirements
Operating temperature 8)19 – 25 ºC (air conditioning recommended)19 – 25 ºC (air conditioning recommended)
Relative humidity 8) 9)< 70% (non-condensing) < 40% (non-condensing) – recommended< 70% (non-condensing) < 40% (non-condensing) – recommended
Electrical requirements100 – 240 V AC, 4.5 A; 50 – 60 Hz100 – 240 V AC, 4.5 A; 50 – 60 Hz
Potencia nominal280 W280 W
Power consumptionStandby: 20 W, Max during wavelength tuning: 100 WStandby: 20 W, Max during wavelength tuning: 100 W
  1. Collinear mode is achieved with an additional external separator box. Dispersion compensation is optional.
  2. Signal output is realized in a non-collinear configuration.
  3. Specified as a percentage of pump power.
  4. Full width at half maximum (FWHM).
  5. Expressed as normalized root mean squared deviation (NRMSD).
  6. Not tunable, optimized for best overall performance. Not simultaneous to OPA output.
  7. Full width at half maximum (FWHM), assuming Gaussian pulse shape.
  8. Specifications are guaranteed for a maximum temperature variation of ± 1 o C and humidity variation of ± 10%
  9. ORPHEUS-MIR output power varies in response to changes in laboratory humidity.